Solid-Phase BBCEAS Measurements at UV wavelength

Authors

  • Noora Sh. Oraha Qas Nouna

Keywords:

Broadband cavity enhanced absorption spectroscopy; BBCEAS, Thin films; Drop coating method; Absorption spectroscopy; Ultraviolet emitted diode; UV-LED.

Abstract

Broadband Cavity Enhanced Absorption Spectroscopy (BBCEAS) is
employed in this study as a sensitive technique for measuring the
absorption of thin films deposited onto a glass substrate. A thin film of the
analytic Rhodamine B was deposited on glass microscope coverslips. Drop
coating method was used to deposit a thin film over the microscope
coverslips. The most sensitive measurements as determined by the
minimum detectable loss per pass (?Fmin(t)) at 397 nm and the R ? 0.99
ultra violet mirror set. A value of ?Fmin(t) = 5.2×10-6 was obtained.

Downloads

Published

2020-09-20

How to Cite

Noora Sh. Oraha Qas Nouna. (2020). Solid-Phase BBCEAS Measurements at UV wavelength. Mustansiriyah Journal for Sciences and Education, 17(5), 197–206. Retrieved from https://edumag.uomustansiriyah.edu.iq/index.php/mjse/article/view/66

Issue

Section

Research Article